www.microscan.ru    
 





Preventive inspection is spent two times a year.

Routine maintenance is at the same time with preventive inspection of the system and includes following works:

  • Cleaning of all optical elements of beam delivery system, measurement of transmission coefficients of optical elements transmission (conclusions about their replacement) and its adjustment.
  • Adjustment of targeting and focusing lasers, adjustment of tracking system videocamera.
  • Test of galvanoscanner performance and matching of it’s co-ordinates system to given parameters.
  • Measurement of parameters of radiation distribution.
  • Measurement of working energy in operational mode.
  • Checking, cleaning (replacement if it need) the mirrors of the laser resonator and its adjustment.
  • Replacement of halogen filter.
  • Calibration of energy of the laser and setting it’s internal parameters.
  • Test of vacuum pump performance.
  • Performance check of high-voltage thyratron and its consumption systems.
  • Performance check gas vacuum system, electromagnetic valves and gas reducer.

Rus  Eng
 
 
   
Email
Troitsk, Moscow Region, 142190, Russia
Tel: +7 (4967) 51-0211
Fax: +7 (4967) 51-0216